Flexure guides are free of maintenance, friction, and wear, and do not require lubrication. Their stiffness allows high load capacity and they are insensitive to shock and vibration. They are 100 % vacuum compatible and work in a wide temperature range.
The patented PICMA® piezo actuators are all-ceramic insulated. This protects them against humidity and failure resulting from an increase in leakage current. PICMA® actuators offer an up to ten times longer lifetime than conventional polymer-insulated actuators. 100 billion cycles without a single failure are proven.
In a parallel-kinematic multi-axis system, all actuators act on a common platform. The minimum mass inertia and the identical design of all axes allow fast, dynamic, and nevertheless precision motion.
S-325.30L | S-325.3SL | S-325.3SD | Unit | Tolerance | |
|---|---|---|---|---|---|
Active axes | Z, θX, θY | Z, θX, θY | Z, θX, θY | ||
Motion and positioning | |||||
Integrated sensor | – | SGS | SGS | ||
Travel range in Z at 0 to 100 V, open loop | 30 | 30 | 30 | µm | +20 % / -0 % |
Tip/tilt angle at 0 to 100 V, open loop | 5 | 5 | 5 | mrad | +20 % / -0 % |
Travel range in Z, closed loop | – | 30 | 30 | µm | |
Tip / tilt angle in θX, θY, closed loop | – | 4 | 4 | mrad | |
Resolution in Z, open loop | 0.5 | 0.5 | 0.5 | nm | typ. |
Resolution in θX, θY, open loop | 0.05 | 0.05 | 0.05 | µrad | typ. |
Resolution in Z, closed loop | – | 0.6 | 0.6 | nm | typ. |
Closed-loop resolution in θX, θY | – | 0.1 | 0.1 | µrad | typ. |
Mechanical properties | |||||
Resonant frequency in Z, no load | 2 | 2 | 2 | kHz | ±20 % |
Resonant frequency, under load (with 25 mm × 8 mm glass mirror) | 1 | 1 | 1 | kHz | ±20 % |
Distance of pivot point to platform surface | 6 | 6 | 6 | mm | ±0.5 mm |
Platform moment of inertia | 515 | 515 | 515 | g × mm² | ±20 % |
Drive properties | |||||
Ceramic type | PICMA® P-885 | PICMA® P-885 | PICMA® P-885 | ||
Electrical capacitance | 3 × 3.1 | 3 × 3.1 | 3 × 3.1 | µF | ±20 % |
Miscellaneous | |||||
Operating temperature range | -20 to 80 | -20 to 80 | -20 to 80 | °C | |
Housing material | Aluminum | Aluminum | Aluminum | ||
Mass | 0.065 | 0.065 | 0.065 | kg | ±5 % |
Cable length | 2 | 2 | 2 | m | +100 mm / -0 mm |
Sensor / voltage connection | LEMO | LEMO | D-sub 25 (m) | ||
Recommended electronics | E-610, E-616, E-663, E-727 | E-610, E-616, E-663, E-727 | E-610, E-616, E-663, E-727 |
The resolution of the system is limited only by the noise of the amplifier and the measuring technology because PI piezo nanopositioning systems are free of friction.
For maximum tilt range, all three piezo actuators must be biased at 50 V. Due to the parallel-kinematics design, linear travel and tip/tilt angle are interdependent. The specified values are the maximum for pure linear respectively tilt motion.
All specifications based on room temperature (22 °C ±3 °C).